Installation; Setting Up Illuminator - Sunoptic Surgical Titan x350 Mode D'emploi

Lampe à xénon avec contrôle numérique d'intensité
Masquer les pouces Voir aussi pour Titan x350:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43

5. INSTALLATION

5.1 SETTING UP THE ILLUMINATOR
Place the illuminator on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the illuminator may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable gases.
This illuminator pulls cooling air into its cabinet through a filtered grille in the
bottom panel underneath the front portion and exhausts warm air from the fan in
the rear.
DO NOT block the space in front or below the front of the illuminator or impede
exhaust air flow behind or above the cabinet.
A clear space minimum of 5"/12.7cm behind and above the cabinet is required.
The illuminator should not be placed where its exhaust will influence other devices,
nor where exhaust from other devices will influence the illuminator.
Ensure the rear power switch is in OFF position.
Ensure that the lamp is properly positioned for use. Open the lamp compartment door.
Move the lamp positioning lever (found to the right of the lamp) clockwise, from horizontal
to vertical. As the positioning cones disengage the lamp should drop about 3mm. Move
the lamp positioning lever counterclockwise to reengage the positioning cones and
observe that the lamp rises about 3mm as it returns to operating position.
Close the lamp compartment door.
Connect AC power cord to the power inlet located on the rear panel of the illuminator.
Ensure the power cord is completely seated into the power inlet and engage the cord
retainer if so equipped.
Use only power cords provided with the illuminator.
Plug the AC power cord into a wall outlet using the three (3) prong plug supplied
CAUTION
with theunit.
To prevent electric shock, connect power cords of peripheral equipment
through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the transformer
power ratings.Ensure that the power cord is connected to the main power with
three prong plug (USA use UL2601-1 rated isolation transformers and/or power
strips only).
Nearby radio or TV units may subject this equipment to radio interference.
To avoid adverse electromagnetic effects, DO NOT operate this equipment
CAUTION
near RF energy equipment.
LIT-218 Sunoptic Surgical
Rev. C
®
Page 9 of 170
(English)

Publicité

Table des Matières
loading

Table des Matières