Sunoptic Surgical Titan x350 Mode D'emploi page 12

Lampe à xénon avec contrôle numérique d'intensité
Masquer les pouces Voir aussi pour Titan x350:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43
6.2.1.4
Installed lamp does not match the intended power (watts). Tx350 units require a 300-watt lamp.
6.2.1.5
Lamp has been used more than 1000 hours. Install new SSX0350 lamp or one used
less than 1000 hours.The main screen always displays lamp hours, and notices are
displayed when a lamp is approaching the end of its useful lamp life. This screen is
displayed when the lamp has exceeded its usable lifetime.
6.2.1.6
In the event of a fan problem the unit will turn off the lamp and display this screen.
Turn off and restart the illuminator. If the fault persists the unit needs servicing
LIT-218 Sunoptic Surgical
Rev. C
INCORRECT WATTAGE LAMP INSTALLED
Install correct SSX0350 lamp.
LAMP HAS EXCEEDED USEABLE LIFE
FAN FAULT
®
Page 12 of 170
(English)

Publicité

Table des Matières
loading

Table des Matières