Battery And Fuse Replacement - Amprobe AM-510 Mode D'emploi

Masquer les pouces Voir aussi pour AM-510:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Except for the replacement of the battery, repair of the meter should be
performed only by a Factory Authorized Service Center or by other qualified
instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent and
water. Apply sparingly with a soft cloth and allow to dry completely before
using. Do not use aromatic hydrocarbons, Gasoline or chlorinated solvents for
cleaning.

BATTERY AND FUSE REPLACEMENT

X�
WARNING
To avoid shock, injury, or damage to the Meter:
Disconnect test leads before opening case.
Use ONLY fuses with the amperage, interrupt, voltage, and speed ratings
specified.
Replacing BATTERY follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the battery cover and open the battery cover
4. Remove the batteries and replace with one 9V (6F22) or equivalent. The
battery cover provides the correct polarity fitting construction design.
Install the battery in the battery cover.
5. Put the battery cover back and re-fasten the screw.
Battery: 9V (6F22) Battery or equivalent
Replacing FUSE follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the enclosure and open the enclosure.
4. Remove the broken fuse and replace with new specified fuse.
5. Put the enclosure back and re-fasten the screw.
Fuse ratings:
mA /μA input terminal:
F1 fuse, 0.5A H 660V fast-fuse, 6.3×32mm (AM-510)
F1 fuse, 0.5A H 700V fast-fuse, 6.3×32mm (AM-510-EUR)
18

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Am-510-eur

Table des Matières