Using the footswitch control:
Connect the footswitch control feeding cable with the plug marked as
After the device has been connected, all Leds are still off. When the ON/OFF button is pressed, all Leds are activated at once for 1
second (autotest). At the end of the autotest cycle, the ON/OFF button led will flash. Press the button marked as (---) to perform
suction using the footswitch control and execute intermittence work cycles. Press the button marked as (-) to perform remote
suction using the footswitch control (continuous suction). To stop suction just push on the footswitch control with strength.
Using the footswitch control and the flow deviator: If using equipment fitted with a flow deviator, users may direct suctioned
liquids in any of the two collection tanks provided. Flow deviator comes with two complete suction kits (2 sets of tubes and two
conical junctions).
After the device has been connected, all Leds are still off. When the ON/OFF button is pressed, all Leds are activated at once for 1
second (autotest). At the end of the autotest cycle, the ON/OFF button led will flash. To decide which side to perform the suction
from, press OUT LEFT or OUT RIGHT and the selected button led will show a blue light.
Press the ON/OFF button again to start the suction cycle. If the device is set up for using the flow deviator, ensure the antibacterial
filter has been positioned on both sides.
Connect the footswitch control feeding cable with the plug marked as
Press the button marked as (---) to perform suction using the footswitch control and execute intermittence work cycles.
Press the button marked as (-) to perform remote suction using the footswitch control (continuous suction). To stop suction just
push on the footswitch control with strength. Press the ON/OFF button to stop the medical equipment.
Before removing the feeding plug, ensure autotest has been performed on the panel.
MAKE SURE THAT CHILDREN AND/OR MENTALLY ILL PEOPLE DO NOT USE THE DEVICE WITHOUT ADULT
SURVEILLANCE
ALWAYS PLACE THE DEVICE IN POSITIONS FOR EASY DISCONNECTION.