Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Subject
Display:
Software and
Electronics
Controls:
Database:
3 x 200 V / 50 Hz
Power Supply
3 x 200-210V / 60 Hz
3 x 220-230 V / 50 Hz
3 x 220-240 V / 60 Hz
3 x 380-415 V / 50 Hz
3 x 460-480 V / 60 Hz
Mains Cable
Recommend-
ation
3 x 200 V / 50 Hz
3 x 200-210 V / 60 Hz
3 x 220-230 V / 50 Hz
3 x 220-240 V / 60 Hz
3 x 380-415 V / 50 Hz
3 x 380-415 V / 60 Hz
3 x 460-480 V / 60 Hz
Important:
Local standards may overrule the recommendations for the main supply cable.
If necessary, please contact a qualified electrician to verify which option is suitable for the
local installation setup.
type A, 30 mA (or higher) is recommended.
Residual
Current Circuit
Breaker
Main motor:
Electrical Data
Sample mover motor:
Carrousel motor:
High pressure water pump:
Recirculation pump for cooling
water:
min 1.0 bar, max. 10.0 bar
Water Supply
min. 5.0 bar, max. 9.9 bar
Air Supply
Prepamatic-2
Instruction Manual

5. Technical Data

Specifications
128 x 240 dots (16 x 40 characters)
touch pad
10 Struers methods
70 user definable methods
max. cont. load 7.3 A
max. cont. load 7.3 A
max. cont. load 7.3 A
max. cont. load 7.3 A
max. cont. load 3.9 A
max. cont. load 3.9 A
Min.
Fuse
size
40 A
40 A
40 A
40 A
32 A
30 A
30 A
750 / 120 W
30 W
20 W
180 W
140 W
Consumption: approx 1.5 l/cleaning
max. 60 l/min, average 15 l/min
Minimum cable size
Max.
@ Min. fuse
Fuse
size
3x4.0mm² + PE
50 A
3xAWG8 + PE
50 A
3x2.5mm² + PE
50 A
3xAWG10 + PE
50 A
3x1.5mm² + PE
50 A
3xAWG14 + PE
50 A
3xAWG14 + PE
50 A
Minimum cable size
@ Max. fuse
3x4.0mm² + PE
3xAWG8 + PE
3x4mm² + PE
3xAWG10 + PE
3x4mm² + PE
3xAWG10 + PE
3xAWG10 + PE
47