Error
Error Message
Dresser movement error.
Stone at minimum, new stone
required.
Specimen holder moving while
carrousel is moving.
Specimen holder moving error.
Low water pressure. Check
water supply.
Timeout in removal mode.
Reduce stock removal or
increase force.
Removal error, force too low.
Reduce stock removal or
increase force.
Carrousel blocked. Remove a
possible hindrance.
Water is low in recirculation
tank.
Stone cover open.
44
Prepamatic-2
Instruction Manual

3. Trouble-Shooting

Explanation
Check air.
The Specimen holder cannot
move up or down.
Action
Check the compressed air
supply and the free movement
of the dresser.
Replace the grinding stone.
Call technical service
Check that the specimen holder
head moves freely upwards. If
necessary, lubricate spindle
through the lubricating nipple in
the dresser room.
If that does not help, call
technical service.
Check that the water inlet and
all hoses are connected
correctly and that the water
pressure is sufficient. If the
water supply has been
interrupted, there might be air in
the system. Re-start 2 - 3 times
by pressing F1:Continue.
Start the process with a higher
force or reduce the total amount
of material to be removed.
Start the process with a higher
force or reduce the total amount
of material to be removed.
Check that the carrousel moves
freely and remove a possible
hindrance.
Refill the Recirculation Cooling
Unit.
Replace the stone cover.