Seca 465 Mode D'emploi Et Garantie page 71

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 77
Fault
Error LED flashing
Error LED flashing
Error LED flashing
"MSG For Children only"
message appears on
the printout
"MSG image not loaded
to printer" message
appears on the printout
Printing does not start.
Incorrect time and date
on printout
Macro executing
- Wait until macro complete.
Power supply too low:
• power supply unit defective
- Replace power supply unit
• printer defective
- Do not continue using printer. Use replacement
device
Power supply too high:
• power supply unit defective
- Replace power supply unit
• printer defective
- Use substitute device
• No reference data are available for the patient age
entered
- Check age entered and correct if necessary
• No logo saved on the SD card
- Select a logo and save it on the SD card - see "Select
your own logo (logo)" on page 61
• It was not possible to transmit measurement results to
the printer.
- Ensure that the printer is switched on.
• Printer not integrated in wireless network.
- Switch off the scale or height measuring device
- De-energize the printer and its wireless module
- Set up the wireless network again.
• Nearby HF equipment (e.g. mobile phones) is
interfering with reception.
- Make sure that HF equipment is kept at least 1 metre
away from transmitters and receivers in the seca
wireless network.
• Date and time not set
- Set date and time using a seca 360° wireless scale
or height measuring device (see instruction manual
for the
- seca 360° wireless device used).
• Internal battery of wireless module exhausted.
- Contact seca Service.
Cause/Remedy
What should I do if ...? • 71
English

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

466

Table des Matières