Owner's Guide
according to the products directions.
Suspend the beaker in the tank by the
use of the beaker holder and place
the beaker positioning ring around the
beaker so that the bottom 2/3 will be
suspended in the main bath.
C. Instrument Draining
The UC125H is equipped with a bezel
surrounding the tank. The bezel drains
condensation back into the tank
keeping the work area around the unit
dry. The bezel also acts as a draining
stand for the basket. Simply position
the corner of the basket on two of
the draining tabs (Fig. 4) at opposite
corners of the tank and allow the
solution to drain off instruments.
IX. Accessory Use Directions
A. Full Size FingerGuard™ BioSonic
UC151XD Instrument Basket cleans
large volumes of instruments and
increases instrument processing
efficiency.
Use Directions:
Place instruments across ridges in
bottom of basket. Place basket in
tank and clean for desired time.
After cleaning rinse basket and
instruments thoroughly under
running water.
B. Three-Quarter FingerGuard: The
UC152XD Basket provides you with
the versatility of cleaning a basket load
of instruments and beaker cleaning
for prostheses (or other small items)
simultaneously.
Use Directions:
Place Basket into UC125H (Fig. 5).
Place instruments across ridges in
bottom of basket. Place basket in tank
and clean for desired time.
C. Beaker Holder
Place three UC153 Beaker Holders
on the bezel. Place positioning rings
around the beaker so that the bottom
2/3 of the beaker is suspended in the
main bath. Clean for desired time.
X. Environmental Conditions
• Intended for indoor use.
• Maximum altitude 2000 m.
•
40°C (104°F)
5°C (41°F)
•
80%@31°C (88°F)
50%@40°C (104°F)
• Main supply voltage fluctuations not to
exceed ± 10% of the nominal voltage.
• Transient over voltages category II.
• Pollution degree 2.
• Class I equipment.
XI.
Cautions and Warnings
NOTE: Failure to comply with these
Caution and Warning statements
may result in injury to personnel or
equipment damage.
Do not operate the unit with-
out sufficient solution in the
tank. Do not use water alone
as the carrier bath. Water is
not a satisfactory coupling
agent unless a wetting agent
is used. For best results use
BioSonic
General
BioSonic Germicidal Cleaner or
BioSonic Enzymatic Ultrasonic
Cleaning solutions.
Do not place objects directly
on bottom of the tank because
they may block cavitation and
cause buildup of heat in the
CAUTION!
transducer
causing
ture failure. Use only BioSonic
UC125H accessory baskets,
cassette racks or beakers to
hold items.
Do not restrict airflow. Provide
adequate ventilation by main-
taining at least 2 inches clear-
ance at the back.
Do not subject unit to shock or
impact.
Do not immerse the unit in
water.
This product is intended for
use only by trained healthcare
professionals.
Use this product in accordance
with the instructions in the
owner's guide.
Unspecified or improper use of
this product may impair safety
protection.
Before cleaning items of con-
siderable value ultrasonically,
consult the item's manufactur-
er for cleaning precautions.
The power off selection from
CAUTION!
the keypad does not turn off
main power to the unit.
This equipment may cause ra-
dio interference or may disrupt
the operation of nearby equip-
ment. It may be necessary to
take mitigation measures such
as re-orienting or relocat-
ing the unit or shielding the
location.
Never drain tank by manually
discarding solution as this may
cause the unit to stop working
properly.
Use only the supplied power
cord or equivalent.
Do not disassemble the unit.
There are no user serviceable
Purpose,
parts inside.
Keep unit covered to reduce
airborne contaminants.
Do not place fingers or hands
in the ultrasonic tank while in
operation.
WARNING!
Operation for long periods of
time will cause the cleaning
prema-
solution to get very hot. To
avoid scalding, do not place
hands in the solution or spill
any onto you. Be careful when
removing an item from the
tank as it may also be hot.
5