7 Maintenance, Cleaning, Diagnosis

This chapter describes the user maintenance and cleaning
procedures for Gnathometer CAD. Only those tasks are listed
that may be performed by dental professionals. All other
tasks must be performed by qualified service personnel at a
certified Ivoclar Vivadent Service Centre.

7.1 Monitoring and maintenance

The time for these maintenance procedures depends on the
frequency of use and the working habits of the user. For that
reason, the recommended times are only approximations.

7.2 Cleaning

Before disinfection, the Gnathometer must be completely
disassembled.
Item
Frequency
Registration
After every
components and
use
small parts
(metal)
Fixing plates
After every
(plastic)
use
Bite rim supports
After every
(plastic)
use

7.3 Maintenance

Sterilize and/or disinfect the components before the first use
and after every use and/or after contact with saliva or blood.
Cleaning
material
Sterilization in
the autoclave
according to the
valid standards
or regulations
Wipe disinfection
using a spray
disinfectant
according to the
valid standards
Wipe disinfection
using a spray
disinfectant
according to the
valid standards
Avoid any contact with strong acids and solvents
(e.g. MMA, acetone) to prevent damage of the
surfaces.
Use only disinfectants suitable for stainless steel and
light metals.
Avoid any contact with strong acids, brines or solvents.
Clean in an ultrasonic bath, with steam jets or
sterilization devices.