Using a standard thermal disinfector for baby
bottles (not a microwave oven)
CAUTION
Risk of infection due to inadequate disinfection
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Make sure that the disinfector is clean and operating
properly before every disinfection process.
• Allow the disinfection to continue until the disinfector
switches off automatically or the minimum disinfection
time stated in the instructions for use of the disinfector has
elapsed. Do not switch the device off prematurely.
Use a thermal disinfector with a runtime of at least 6 minutes.
Regarding the performance of the disinfection, the duration of
the disinfection procedure and the quantity of water required,
follow the instructions for use of the disinfector you are using.

4.4 Care of the connection tube

Dry the connection tube after every inhalation session:
• Connect the connection tube to the compressor.
• Switch your compressor on.
• Leave the compressor to work until the air flowing through
the tube has removed all traces of any condensation in the
tube.

4.5 Inspecting

Inspect all product components after each cleaning and
disinfection. Replace any broken, misshapen or seriously
discoloured parts.