Tecno-gaz KYRI Mode D'emploi page 18

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
ENGLISH
 
The maximum allowable weight on the chair is 180 Kg. This value must
not be exceeded.
Do not sit on the tip of the chair when this is tilted upwards.
Do not rest against or sit on the leg guard when it is in table configuration.
The maximum weight capacity is 30 kg.
To avoid the risk of electric shock, connect this unit only to a power
supply network with protective earthing.
Make sure that the voltage of the electrical system is stable and always
remains within the legal range: 230Vac +/- 10% , 50/60 Hz.
Unplug the unit before leaving the surgery.
The unit is not protected against the penetration of liquids.
To avoid contamination, it is advisable to cover all the upholstery with a paper or cotton
sheet to offer protection and prevent promiscuous contact with it, and to replace the cover
after each patient.
The unit is not suitable for use in the presence of a mix of anaesthetic gas that can catch
fire in contact with oxygen or nitrous oxide.
The equipment was designed to ensure continuous operation with intermittent loads. The
recommended operating times and pauses are specified below:
Operation: 1 min / Pause: 10 min.
Tecno-gaz guarantees the safety, reliability and performance of the equipment.
The terms of the guarantee are as follows:
observance of the conditions on the guarantee certificate; the equipment must be used
only in accordance with the instructions in this booklet;
the electrical system in the room where the equipment is installed must be compliant with
the rules in force. Assembly, repairs, additions to the unit, calibration, and all operations in
general that involve opening the guards, must always be carried out by the authorised
technicians.
Attention! Use only original spare parts.
4
 

Publicité

Table des Matières
loading

Table des Matières