Autoclave Use And Applications - Tecno-gaz Europa B EVO Mode D'emploi

Autoclave
Masquer les pouces Voir aussi pour Europa B EVO:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 83
ENGLISH
03
The autoclave is able to sterilize the three types of load provided for by the standard
EN13060, especially:
METAL OR SOLID
MATERIALS
POROUS OBJECTS
HOLLOW OBJECTS
This device has been certified for the sterilization of the following materials *:
Wrapped and unwrapped solid
materials for a maximum load of
Type A/B hollow materials for a
maximum load of
Porous materials for a maximum
load of
* Only for European Countries
The autoclave shall only have to be used for the sterilization of tools and
materials being compatible with the steam sterilization system.
Always make sure that the loads that need to undergo sterilization can
stand the temperatures scheduled for the selected cycle.
8

AUTOCLAVE USE AND APPLICATIONS

Instruments with no cavities and no obstacles to the penetration of
steam
Simple or composite materials that can absorb fluids (fabrics,
gowns, surgical gauzes, dressings, etc)
Materials or devices with cavities, obstructions, etc.
These are subdivided into two types, classified according to the
length and diameter of the cavity. Approximately:
TYPE B: cannulas, tubes or devices with large passages.
TYPE A: turbines, hand pieces and devices with blind or small
holes
EUROPA B EVO
max kg. 5
max kg. 5
max kg. 1,5
EUROPA B EVO 20 / 24
max kg. 7
max kg. 7
max kg. 2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Europa b evo 20Europa b evo 242033s22033clsStga099

Table des Matières