02
USE AND DESTINATION OF USE OF THE AUTOCLAVE
The autoclave is able to sterilize the three types of load provided for by the standard
EN13060, especially:
METAL OR SOLID MATERIALS
Instruments with no cavities and no obstacles to
the penetration of steam
POROUS OBJECTS
Simple or composite materials that can absorb
fluids (fabrics, gowns, surgical gauzes,
dressings, etc)
HOLLOW OBJECTS
Materials or devices with cavities, obstructions,
etc. These are subdivided into two types,
classified according to the length and diameter
of the cavity.
Approximately:
TYPE B: cannulas, tubes or devices with large
passages.
TYPE A: turbines, hand pieces and devices with
blind or small holes
The charges (Kg) change depending to the type of cycle which is performed. Watch
Fig.B.
*Only for European Countries
The autoclave shall only have to be used for the sterilization of tools and
materials being compatible with the steam sterilization system.
Always make sure that the loads that need to undergo sterilization can
stand the temperatures scheduled for the selected cycle.
4
ONYX-B 5.0
max kg. 5
max kg. 1,5
max kg. 5
ONYX-B 8.0
max kg. 8
max kg. 2
max kg. 8