6. Place the autoclave at a height that allows the user to inspect and easily clean all parts
of the sterilization chamber.
7. Open the door of the autoclave and remove from inside the sterilization chamber all
packages which contain the single accessories.
8. Leave inside the sterilization chamber only the tray carrier with the trays. All other
accessories should be positioned in a separate space available to operators.
9. Never place on the autoclave newspapers, trays, liquid containers, etc...
10. Do not lean on the door when it is open.
11. Leave a space of at least 5 cm in the rear of the autoclave, using the spacer
(FIG.1–pos.
0
12. Connect the overflow connectors to the special drains with the pipes provided
(chapter 5)
13. Always make sure the electrical system to which the autoclave is to be connected is in
conformity with the legislation in force and sized to suit the specifications of the said
appliance.
14. Take the power supply cable supplied and plug the plug on the rear panel of the
autoclave into the female socket (FIG.1 –pos.
15. Connect the electric plug to the system and make sure the power supply is 230 Vac -
2000 Va minimum
Do not make the connections using extensions, reducers or adapters as
this could create micro outages with consequent generation of alarm
NOTE:
signals.
16. Turn on the autoclave by pushing the general switch (FIG.2–pos.
door to the autoclave. Wait a few seconds, there will be two beeps which indicate the
reaching of relative parameters for automatic atmospheric alignment, at the same time
the display will indicate "open door".
Never select a control before hearing the two acoustic signals: the
NOTE:
autoclave will not accept the selected program
17. Load the pure water tank following the indications in the section entitled usage
instructions (chapter 8).
Successively run the test cycle, always following the indications expressed in the
section usage instructions (chapter 8).
18. If the autoclave is installed with the demineralizer system, follow the instructions
contained in the packaging of this device and the demineralizer section (chapter 16).
18
/ FIG.18), and sides of the unit to ensure the ventilation required
6
)
17
) and open the