becomes smooth and shiny.
- The handpiece must be cleaned properly and
regularly as mentioned in the user's manual.
Using the device:
- Do not use the device if it appears to be damaged
or faulty.
- Before unplugging the mains power cable, switch
the device off.
- To unplug the mains power cable, pull the plug on
the end of the cable whilst holding the wall socket
with your other hand.
- Never use irrigating solution containers that are
not designed to be suspended from the bracket
provided.
- The device must only be used with bottles or pouches
of sterile water or physiological saline solution.
- The capacity of the irrigating solution container
used shall not exceed one litre.
- If you are not using or will be away from the device
for a period of time, disconnect the device from the
electrical supply network.
- Do not move the device during use.
- Do not apply excessive pressure to the screen.
Environment:
- Do not cover the device or obstruct the air inlets.
- Do not immerse and do not use outside.
- Do not tilt the device at an angle of more than 5°.
- Do not place the device near a heat source.
- Make sure that the cables do not prevent people
from moving about freely.
- The device shall be stored in its original packaging,
in a suitable place, which does not present a hazard
for people.
- The device is not designed to be used in a medical
operating room.
- The device is not designed to operate in the presence
of anaesthetic gases or any other Á ammable gas.
- Do not expose the device to water mist or spray.
- Any condensation inside an electrical device is
potentially dangerous. - If the device is to be moved
from a cold environment to a hot one, it must not be
used until it has reached room temperature.
- The device is not designed to operate in the vicinity
of ionizing radiation.
6
- Do not insert metal objects into the device to
prevent any risk of electric shock, short-circuit or
the emission of hazardous substances.
Servicing:
- Before and after each use, it is vital that you
disinfect your device using products recommended
by SATELEC.
- Prior to each intervention, it is imperative that you
use clean, disinfected and sterilized accessories.
Accessories:
- The device may only be À tted with accessories
designed or recommended by SATELEC that are
perfectly suited to the purpose.
- Use of accessories from other manufacturers is a
potential hazard for you or your patients.
Repair:
- Warning: Do not repair or modify the device without
the prior authorization of SATELEC.
- Warning: If the device is modiÀ ed or repaired,
speciÀ c inspections and checks shall be performed
to ensure that the device can still be used safely.
- In the event of an anomaly, contact the supplier of
your device. Do not contact any other repairer who
could render your device hazardous for you and your
patients.
In the event of doubt, contact an approved dealer or
the Customer service department:
- www.acteongroup.com
- E-mail: satelec@acteongroup.com.
III - TECHNICAL DATA
3.1 PHYSICAL DESCRIPTION
The device comprises 5 main parts:
- A console (Fig. 1-1).
- A multifunction pedal (Fig. 4).
- A P
S
or P
IEZOTOME
OLO
cable (Fig. 1-2).
- A bracket for irrigating solution (Fig. 1-3).
- A mains power cable with ground (Fig. 1-4).
S
L
handpiece and
IEZOTOME
OLO
ED