Filling The Measuring Cell Using The Built-In Sample Pump; Filling The Measuring Cell Using An External Syringe - Mettler Toledo Densito 30PX Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Measurement (meas)
6.2

Filling the measuring cell using the built-in sample pump

- Ensure that the sample tube is in the sample.
- Press the fill button (12) slowly to fill the measuring cell.
- Ensure that there are no air bubbles in the measuring cell.
6.3

Filling the measuring cell using an external syringe

11
- Press the drain button (3).
- Slide up the lock button (11).
- Remove the screw stopper (6) with a coin.
- Screw in the adapter for the external syringe.
- Hold the sample tube over a waste container.
- Slowly inject the sample into the measuring cell.
- Ensure that there are no air bubbles in the measuring cell.
- Do not remove the syringe while measuring.
14
3
6
METTLER TOLEDO Densito 30PX

Publicité

Table des Matières
loading

Table des Matières