The ground resistivity is calculated by the Wenner method using the following formula :
Schlumberger method:
The ground resistivity is calculated by the Schlumberger method using the following formula:
4.5.2

PRE-MEASUREMENT SETTINGS

The resistivity measurement can only be performed in MANUAL mode
!
Before the measurement, use the
- select the method : Wenner or Schlumberger,
- select the measurement frequency (55, 92, 110, 119, 128Hz, or USr
- select the voltage applied (16 or 32 V).
(1)
The frequency "USr" is defined by the user for between 41 and 128 Hz using the
The distances required for the ρ calculation are programmed using the "DISTANCE"
nd
nd
nd
2
2
2
+
+
CONFIG
CONFIG
CONFIG
key :
DISTANCE
DISTANCE
- one distance only ,"d", for the Wenner method,
- two distances, "A" and "d", for the Schlumberger method.
The values are chosen using the
Notes :
the distances "A" and "d" cannot be modified during the measurement ;
during the measurement, you can use the
frequency, between 41 and 128 Hz only.
ρ
= 2 x Π x d x R
S
S-ES
ρ
= (P x (d² - A²/4)/A) x R
S
S-ES
CONFIG
CONFIG
CONFIG
and
keys to successively :
DISTANCE
DISTANCE
keys
, before or after the measurement.
CONFIG
CONFIG
CONFIG
and
DISTANCE
DISTANCE
49
(1)
,
keys.
keys to modify the measurement