BOND
Aspirating Probe Cleaning System
™
Catalog No: CS9100
Intended Use
The BOND Aspirating Probe Cleaning System is intended for automated cleaning of aspirating probes on BOND instruments, using the
aspirating probe cleaning protocol installed in BOND systems (includes Leica BOND-MAX system and Leica BOND-III system).
Summary and Explanation
Routine use of the BOND Aspirating Probe Cleaning System keeps BOND system aspirating probes free of DAB residue.
The BOND Aspirating Probe Cleaning System is a tray of three cleaning reagents that is loaded onto the BOND system like any other set
of reagents. A 20-minute predefined cleaning protocol mixes reagents from the system in the BOND mixing station then uses a sequence
of aspirations and dispenses in the station to thoroughly clean the bottom 2–3 cm of the probe. As a final step the probe is rinsed with
BOND Wash to remove any cleaning reagent residue.
Use of the cleaning system does not remove the need for regular, manual cleaning of the probe. Leica Biosystems recommends that
users wipe the external surface of the probe with a soft cloth dipped in 70% alcohol once a week.
Reagents Provided
Reagents sufficient for 15 cleans:
1. Cleaning Component A (3.75 mL) proprietary cleaning reagent
2. Cleaning Component B (3.75 mL) <5% sulphuric acid
3. Cleaning Component C (3.75 mL) proprietary cleaning reagent
Dilution and Mixing
The BOND Aspirating Probe Cleaning System is optimized for use on the BOND system. No manual reconstitution, mixing, dilution, or
titration is required. BOND mixes components to the correct concentrations on the instrument when cleaning is initiated.
Materials Required but Not Provided
No additional materials are required to use the BOND Aspirating Probe Cleaning System.
Storage and Stability
Store at 2–8 °C. The product is stable under these conditions up to the expiry date indicated on the container and tray handle labels.
There are no obvious signs to indicate instability of this product.
Return to 2–8 °C immediately after use.
Storage conditions other than those specified above must be verified by the user.
Precautions
•
This product is intended for in vitro diagnostic use.
CLEANING
COMPONENT A
Contains Potassium
Permanganate (<1%).
GHS09: Environmental.
CLEANING
COMPONENT B
Contains Sulphuric Acid
(<10%).
GHS07: Exclamation
mark.
Signal words: Warning.
CLEANING
COMPONENT C
Contains Oxalic Acid
(<10%).
GHS07: Exclamation
mark.
Signal words: Warning.
H411: Toxic to aquatic life with long
lasting effects.
H315: Causes skin irritation.
H319: Causes serious eye irritation.
H302 + H312: Harmful if swallowed
or in contact with skin.
P273: Avoid release to the environment.
P391: Collect spillage.
P264: Wash hands thoroughly after handling.
P280: Wear protective gloves/protective clothing/eye protection/
face protection.
P302+352: IF ON SKIN: Wash with plenty of soap and water.
P332+313: If skin irritation occurs: Get medical advice/attention.
P362: Take off contaminated clothing and wash before reuse.
P305+351+338: IF IN EYES: Rinse cautiously with water for
several minutes. Remove contact lenses, if present and easy to
do. Continue rinsing.
P337+313: If eye irritation persists: Get medical advice/attention.
P264: Wash hands thoroughly after handling.
P270: Do not eat, drink or smoke when using this product.
P280: Wear protective gloves/protective clothing/eye protection/
face protection.
P301 + P312: IF SWALLOWED: Call a POISON CENTRE or
doctor if you feel unwell.
P330: Rinse mouth.
P302 + P352: IF ON SKIN: Wash with plenty of soap and water.
P312: Call a POISON CENTRE or doctor if you feel unwell.
P363: Wash contaminated clothing before reuse.
P501: Dispose of contents/container to hazardous or special
waste collection point.
CS9100