Oil Service And Test Run - KLS Martin group Angulus2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 86
Instructions for Use
Angulus2 Angled Screwdriver
6.3

Oil Service and Test Run

NOTICE
Risk of material damage!
Possible damage to products in the case of a failure to comply with oil care.
Oil Service
Use only W&H Service Oil F1.
Follow the instruction for use on the oil spray can.
Under all circumstances perform a care cycle after each cleaning of the inside and before
each sterilization.
Test run following oil service
Position the angled screwdriver with the head pointing downwards.
Operate the angled screw driver to expel excess oil.
Revision 04
53

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

90-601-58-10

Table des Matières