Change Canister/Tubing Set - Medela Invia Motion Mode D'emploi

Masquer les pouces Voir aussi pour Invia Motion:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

When the battery is fully charged, it should last approximately 10
hours. If the Invia Motion NPWT pump is running continuously due
to an air leakage, the battery may need to be recharged earlier
(after approximately 4 hours).
CAUTION
NPWT therapy must be continued while charging the Invia
Motion NPWT pump's battery. Do not turn the Invia Motion
NPWT pump off while charging the battery.
Always charge the battery overnight. Make sure the battery
compartment on the back of the Invia Motion NPWT pump is
closed. Do not remove the battery from the Invia Motion NPWT
pump.

Change canister/tubing set

To be used by healthcare professionals or adequately trained
caregivers only.
Change the canister minimum once a week or when it appears full,
when the canister full signal sounds or when the canister full
symbol
is displayed.
1. Wash hands and prepare clean paper towel.
2. Remove the Invia Motion NPWT pump from the carrying case.
Switch the Invia Motion NPWT pump off by pressing the
button and holding for 3 seconds. Press only once.