3. Position the clamp next to the quick-connector and close the
clamp on the pump tubing and close the clamp on the dressing
tubing.
Clamp open
4. Disconnect the Invia Motion NPWT pump tubing from the dres-
sing tubing by pressing on the sides of the quick-connector.
Have a paper towel or tissue available to cover the tip of the
used tubing to protect against wound fluid leakage.
5. Detach the full canister/tubing set by rotating the canister to the
left.
6. Wash hands.
7. Obtain a new Invia Motion canister/tubing set and open the
packaging.
CAUTION
Do not use an Invia Motion canister/tubing set if the sterile
packaging is damaged or has expired.
Clamp closed
17