Using oxygen tubes (only for navy)

Slowly open the valve for the compressed gas supply.
Provide the T-piece (1) with two pieces of hose, each 5cm long,
and insert in the Dräger tube holder H
Disconnect the rubber cap on the connection tube and break off
both tips.
Break off both tips on the oxygen tube.
Connect the oxygen tube (2) to a piece of hose about 5cm long.
Observe the green marking and arrow on the oxygen tube.
Connect the entry side of the oxygen tube to the T-piece.
Connect the entry side of the connection tube (4) to the hose (3).
Connect the output side of the connection tube to the accuro gas
detection pump. Note the associated Instructions for Use.
Press the accuro gas detection pump (5) together as far as it will
go once, then loosen it and wait for the end of the stroke.
Close the valve for the compressed gas supply, immediately
remove the oxygen tube from the holder, and analyse the results
in accordance with the Instructions for Use.
The oxygen tube heats up during the measurement. Do not use in
an explosive area, as this is an explosion hazard.
Place the rubber cap on the connection tube. The connection tube
can be used four times.
O.
2
CAUTION
Oil
CO
CO H
O
4
2
2
3
2
1
5
27