After cleaning/sterilising, the turbine should be rinsed in distilled water and left to
dry.
Do not use objects, such as pens or cotton buds, to clean or dry the turbine, as
this may cause irreparable physical damage.
To replace the turbine hold the unit and the turbine swirl plate facing you, put
the turbine back into the turbine housing then push until the turbine is fully
located, such that the back of the turbine is completely level with the back of the
unit.
The PulmoLife moulding can be wiped clean using a damp cloth or surgical wipe.
The unit should not require recalibration. However, as a confirmation of
the correct functioning of the unit, the calibration can be checked using a
3-litre calibration syringe.
To check the calibration hold down the On
together, and only release when 'Ch' is displayed.
When the blow symbol
into the unit. The value will be displayed.
If the unit detects an error the blow symbol and calibration value will flash. An
acceptable calibration check is within +/-3% of the 3-litres entered (so between
2.91 and 3.09 litres).
PLEASE NOTE: Calibration check failure is unusual but is most often due to
malfunction of the turbine, either because it is physically damaged or dirty. Clean
the turbine (see cleaning) and recheck the calibration. If the unit still fails
calibration check, replace with a new turbine (see accessories for order number).
This unit requires no routine servicing.