Connect the power cord with the power supply and the
power pack. Make sure that the power cord is easily
accessible at all times and and can easily be disconnect-
ed from the power supply. The charging base will briefly
display "Bluephase G4" as ticker text on the screen.
Handpiece
Unpack the handpiece from the packaging and detach
the light guide by pulling it out. Then clean the
handpiece and light guide (see Maintenance and
Cleaning). After cleaning, reinsert the light guide.
For reasons of hygiene, we recommend using a
disposable protective sleeve for each patient
(see Maintenance and Cleaning). Make sure to fit the
protective sleeve snugly to the light guide. Then, attach
the anti-glare cone or anti-glare shield to the light
guide.
Battery
We recommend fully charging the battery before the
first use! If the battery is fully charged, it features a
curing capacity of approximately 20 minutes. Slide the
battery straight into the handpiece until you hear and
feel it click into place.
Gently place the handpiece in the corresponding rest in
the charging base without using any force. If a hygiene
sleeve is used, please remove it before you charge the
battery. If possible, use the light always with a fully
charged battery. This will prolong the service life. It is
therefore recommended to place the handpiece into
the charging base after each patient. If the battery is
fully discharged, the charging time is 2 hours.
Since the battery is an expendable part, it has to be
replaced after its typical life cycle has expired after
approximately 2.5 years. See battery label for the age
of the battery.
Battery: charging status
The current charging status is displayed on the
handpiece as described on page 6.
11