Operating Principle Of Charger Gt203/Li-Ion Battery Gt201; Charging Principle; Charging Time; Preparation And Setup - B.Braun Aesculap FAV5 CL/FORTIS Mode D'emploi

Masquer les pouces Voir aussi pour Aesculap FAV5 CL/FORTIS:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
2.6 Operating principle of charger
GT203/Li-ion battery GT201
The charger GT203 is designed for the mains voltage/
frequency range 100 V to 240 V / 50 Hz to 60 Hz.
To make the charger GT203 ready for operation, con-
nect it to mains power by plugging in the mains plug.
The charger GT203 is equipped with to charging bays.
The first charging bay is for charging the battery-pow-
ered clipping machine FAV5 CL/FORTIS, while in the
second charging bay the Li-ion battery is charged.
Each charging bay has an indicator field on the charger
front panel assigned to it.
Charging starts automatically, as soon as a Li-ion bat-
tery is inserted in the charging station.
The charging time in the charging bays depends on the
charge state and capacity of the Li-ion batteries.

2.6.1 Charging principle

The Li-ion batteries are charged by constant current
pulses.
The charge state of the Li-ion battery is monitored
during charging. Monitoring of the charging curve
ensures that the battery is fully charged without over-
charging.
Additionally, the battery temperature and charging
time are monitored as well.
The charger monitors the battery status and indicates
the charge state via two LED indicators.
LED indicators:
• green = Battery fully charged
• red = Battery completely discharged

2.6.2 Charging time

Charging is stopped as soon as the maximum charging
time is reached.
The charging time is approx. 50 min.
3.

Preparation and setup

Non-compliance with the following instructions will
preclude all responsibility and liability in this respect
on the part of Aesculap.
When setting up and operating the product, adhere
to
– national regulations for installation and opera-
tion,
– national regulations on fire and explosion protec-
tion.
– operating advisories according to IEC-/VDE regu-
lations.
Fire and explosion hazard!
Do not operate the product in
explosion-hazard areas.
DANGER
Malfunction of, or damage to the
product caused by improper setup!
Do not expose the product to
CAUTION
direct sunlight or moisture.
Do not cover the ventilation
slots of the product.
Connect the product to the
specified voltage of 100 V to
240 V/50 Hz to 60 Hz.
Ensure that the mains power
socket used for supplying the
product is freely accessible.
5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Aesculap suhl fav5 clAesculap suhl fortisGt306

Table des Matières