Resusci Anne DFU 6736C.qxd
Resusci Anne
A A
B B
C C
SkillReporter
Consistent and objective
real time feedback on
performance during practice
helps develop
and reinforce skills.The SkillReporter has a metronome B built
in providing a rate of 80 or 100 strokes/minute, and provides
feedback to the instructor and student to support the student's
initial skill development by means of colour LEDs. In addition the
SkillReporter gives you the possibility to print a report of the
performance (either complete report with ventilation and
compression curves and a statistical report or a statistical
report only).
To use
Turn the SkillReporter on by pressing the "on/off" A key either
on the SkillReporter or on the right side of the manikin torso.
Caution: Make sure the chest skin is properly fastened before tur-
ning on the SkillReporter/manikin. Pressing a second time will turn
the SkillReporter off.The SkillReporter and manikin will turn off
automatically if no activity takes
place for 10 minutes
(60 minutes if the log function is
active). Note:The manikin can
also be used without the
SkillReporter connected.You are
then able to log the training
scenario and later on reconnect
the SkillReporter to print a report.
The light indicators as described below will give instant feedback
on CPR performance.
Light indicators on the SkillReporter
Inflation volume:
Compression depth:
Too fast inflation:
*Measurement tolerance in the manikin: +/- 15% in normal room temperature.
Measurements based on ILCOR guidelines 2005.
6
04-12-06
12:49
Page 6
D D
E E
A C
*
LED(s)
Indication
Limits
yellow
insufficient
<0.5 l
green
sufficient
0.5 l - 0.8 l
red
excessive
>0.8 l
yellow
insufficient
<38 mm (1 1/2")
green
sufficient
38 mm - 51 mm
(1 1/2" - 2")
red
excessive
>51 mm (2")
red (sufficient ventilation < 1 sec. )
The log function
To be able to print a report the CPR data first need to be
logged by activating the log function.The log function can
be activated either on the SkillReporter if connected to the
manikin, or on the manikin.To activate the log function on
the SkillReporter you first have to connect it to the manikin
(the SkillReporter has no internal power source), then press the
"On/Off" button once and after the selftest is complete press
the "LOG" C button.
If you decide to use the manikin without the SkillReporter
attached, press the "On/Off" button at the right side of the torso
and then press the "LOG" C button to activate the log function.
To stop the log function press the "LOG" button a second time.
The CPR data is now logged and can be printed multiple times.
Just connect the SkillReporter and press one of the print but-
tons.The data will not be lost even if the SkillReporter and the
manikin are switched off. When starting a new log, the current
log will be cleared and data cannot be retrieved.
Student name:
Printing a report
Instructor name:
Duration of the session: (min : sec)
Short report D
Ventilations:
Average volume (ml)
The short report format
Average number per min. (#/min)
Minute volume (ml/min)
(statistical data), can only be
Total number (#)
Number correct (#)
printed after the CPR(-D) data
Percent correct (%)
Too much (#)
has been logged. If the "short
Too little (#)
print" button is pressed during
Too fast (#)
logging of a scenario, a sum-
Compr. / Vent. ratio
mary report (statistical data) of
Compressions
Average depth (mm)
the performance will be printed
Average number per min. (#/min)
Average compression rate (#/min)
with data collected up to the
Total number (#)
time of activating the short
Number correct (#)
Percent correct (%)
report. If the "short print" but-
Too deep (#)
Too shallow (#)
ton is pressed after the logging
Wrong hand position (#)
Hand position too low
of a scenario has been com-
Incomplete release (#)
pleted, a summary report of the
Defibrillation (CPR-D models only)
performance will be printed
Time from scenario start
to first shock (min : sec)
with data for the whole session.
Time from call for help
to first shock (min : sec)
Time from arrival of defibrillator
to first shock (min : sec)
Pulse
ResusciAnne
SkillReporter
_______________________
_______________________
____
____
____
____
____
____
____
____
____
____
____
____
____
____
____
____
____
____
____
____
(#)
____
____
____ ____
____ ____
____ ____
Wrong hand position:
yellow
Pressure outside
of acceptable area
(sides and
upper part)
red
Pressure outside
of acceptable area
(lower center)