Trouble-Shooting - Ugin ARTIS NEO Mode D'emploi

Masquer les pouces Voir aussi pour ARTIS NEO:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Instructions for use (rev 290319)

10. TROUBLE-SHOOTING

Type of fault displayed continuous alarm sounds: operator action required
 First stop the alarm by pressing the key
 Depending on the type of fault displayed, consult the list below to determine the cause or
causes of failure and contact your After-Sales Service
 The different types of faults are numbered
o FAILURE 1: THERMOCOUPLE MALFUNCTION
The thermocouple is broken and/or connecting cable (orange)
o FAILURE 2: HEAT MALFUNCTION
Programmed final temperature exceed by + 50°C. It is generally caused by:
- an anomaly in TRIAC on the furnace circuit board
- a new cycle start too near to the previous one
o FAILURE 4: VACUUM PUMP MALFUNCTION
There are two main causes:
- the pump is out of order
- the electric or pneumatic connection is defective or non-existent
Check the cleanness of the tray and the seal (ceramic splinter)
o FAILURE 8: AIR MALFUNCTION
The solenoid valve is clogged or improperly connected to the board (GREEN/GREEN WIRE).
o FAILURE 16-32: PLATFORM MALFUNCTION (UP-DOWN)
The platform is blocked in position, for three possible reasons:
- The back-geared motor is out of order
- The transmission belt is broken
- The electric connection on the board is defective (RED/BLUE WIRE)
- the motor pinion is loose on the motor shaft
 After the correction of the problem, push the button
display. Start the cycle again for verification (just for a test, no-load).
CAUTION: the list given above is only partial, in that certain displays may combine several failures.
Example: FAILURE 22 = FAILURE 16+4+2
Whatever the case, it is essential that you contact your After-Sales Service.
UGIN DENTAIRE
in order to restore the normal
ARTIS NEO
34

Hide quick links:

Publicité

Chapitres

Table des Matières
loading

Table des Matières