I.3
Filter for CO, H2-comp., NO exch-
anging sensors
The measuring instrument must be switched off and the mains unit disconnected from
the mains supply.
1 Place measuring instrument on its face.
2 Loosen screws with a screwdriver, release clip in the direction of the arrow, and
remove service cover.
3 Remove hose connections from sensor.
4 Remove sensor from slot.
.
I.4 Réétalonnage des cellules de mesure
Réglage valeurs, p. 31.
I.5
Nettoyage de la sonde
5 Remove spent filter from sensor.
6 Place new filter on sensor.
Avoid touching the electronics of the sensor.
Observe the markings on the filter and the sensor
7 Insert sensor into slot.
8 Replace hose connections on to sensor.
9 Replace and close service cover (clip must click
in), fix with screws.
10 Reset ppm hour counter (see Display ppm/hour
counter, p. 32.
Séparez la sonde de l'appareil avant le nettoyage.
1 Dégagez le verrouillage de la canne en
actionnant la touche sur la poignée et retirez la
canne.
Pour les sondes à préfiltre :
Dévisser le préfiltre
2 Soufflez les conduits de gaz de la sonde ainsi que
la poignée (cf. illustration). N'utilisez aucune brosse !