H. Environment requirement
Operating Temperature: 5~40°C
Operating Humidity:
Atmospheric pressure:
I. The performance under low perfusion condition
The accuracy of SpO
described above when the modulation amplitude is as low as 0.6%.
J. Resistance to ambient light interference:
The accuracy of SpO2 and PR measurement still meets the specification
described above when the device is tested by SpO2 simulator (Fluke
Biomedical Index 2 series) while setting the emulating interference of sun
light and 50Hz/60Hz fluorescent light.
K. Dimensions: 60 mm (L) × 33 mm (W) × 30 mm (H)
Net Weight: 35g (including battery)
L. Classification:
The type of protection against electric shock: Internally powered
equipment.
The degree of protection against electric shock: Type BF applied parts.
The degree of protection against harmful ingress of liquids: Ordinary
equipment without protection against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B
5. Accessories
A. A lanyard
B. Two batteries
C. A pouch
D. An External SpO
E. A User Manual
F. Installation CD (optional)
Note: The accessories are subject to change. Detailed items and quantity
see the Packing List.
6. Repair and Maintenance
6.1 Oximeter Maintenance
The service life (not a warranty) of this device is 5 years. In order to ensure its
long service life, please pay attention to the use of maintenance.
ENGLISH
30~80%
70~106kPa
and PR measurement still meet the precision
2
Probe (optional)
2