7.2 Lubricating the Lift

WARNING!
Risk of Falling
Hydraulic oil or lubricant on the floor can cause a
fall and injury.
– Wipe excess lubricant from the lift after
lubrication.
– Use a facial tissue to wipe excess oil from
around the hydraulic piston.
– If excess oil is leaking from the hydraulic pump,
contact a dealer or qualified technician for
service.
The Invacare lift is designed for minimum maintenance.
However, a six month check and lubrication should ensure
continued safety and reliability.
Keep lift and slings clean and in good working order. Any
defect should be noted and reported to your Dealer or
Invacare representative as soon as possible.
1551341-E
Refer to the figure for lubrication points. Lubricate all pivot
points with a light grease (waterproof auto lubricant). Wipe
all excess lubricant from lift surface.
1. Hanger Bar
2. Boom Mounting Bracket
3. Boom/Mast Mount
4. Mast Mounting Bracket

7.3 Cleaning the Sling and Lift

Cleaning the Sling
Refer to the washing instructions on the sling and to the
sling manual for cleaning details.
Cleaning and Disinfecting the Lift
CAUTION!
Risk of Damage
Motors, control unit and mounting parts can be
damaged if the lift is cleaned improperly.
– Never use acids, alkaline or solvents for
cleaning the lift.
– Dry the lift carefully after cleaning.
To prevent cross-infection, the hoist must be cleaned and
disinfected after each use.
A soft cloth, dampened with water and a small amount of
mild detergent, is all that is needed to clean the patient lift.
The lift can be cleaned with non-abrasive cleaners.
Never use acids, alkaline or solvents for cleaning the lift. Dry
the lift carefully after cleaning.