5.4
TIME MODE
•
If the remote control display shows °C or °F, press MODE to select TIME
MODE. The remote control display shows 't' in TIME MODE.
•
Press UP or DOWN to adjust the time in steps of 0.1 minute.
•
Press POWER to select the power level. Use the guidelines in section 5.8 to
determine the correct power setting.
•
Press START/STOP to start the heater. The remote control display shows the
time that remains.
•
During heating, the temperature measured by the probe can be displayed for
a couple of seconds by pressing MODE.
•
When the time has elapsed, the heater demagnetises the workpiece, switches
off, and generates an acoustic signal for 10 seconds.
•
Press START/STOP to cancel the acoustic signal and stop the heater.
•
Remove the workpiece with proper handling equipment.
•
The TIH 100m is now ready to heat another workpiece with the same
settings.
5.5
Temperature measurement
When the heater is not operating, the temperature of the workpiece can be
measured by pressing MODE and START/STOP at the same time. The LED on the
START/STOP button flashes during temperature measurement. Press START/
STOP to cancel temperature measurement.
5.6
Change of temperature unit
Press MODE and UP at the same time to switch between °C and °F.
The temperature unit setting remains the same even after disconnection from
mains power.
5.7
Demagnetisation
The workpiece is automatically demagnetised when heating is complete.
Demagnetisation will not occur if the power is interrupted or the main switch is
switched off. To use the TIH 100m for demagnetisation only, select TIME MODE
and set the time to 0.1 minute (6 seconds).
14
SKF TIH 100m