Technical Specification - arjo Flowtron Excel Manuel D'utilisation

Masquer les pouces Voir aussi pour Flowtron Excel:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 78

11. Technical Specification

Pump
Supply Voltage and Frequency:
Power Input:
Size:
Weight:
Case Material:
Plug Fuse Rating:
Pump Fuse Rating:
Degree of protection
against electric shock:
Degree of protection
against liquid ingress:
Mode of operation:
Cycle or Therapy Modes:
Pressure Range:
ENVIRONMENTAL INFORMATION
Condition
Operating
Storage and Transport
(Long Term)
Storage and Transport
(Short Term)
Products should be well packaged and stored in a well-ventilated, dry and non-corrosive environment.
NOTE
If this pump unit is subjected to extreme cold during transit or storage, initial
performance may be impaired. Connect to power supply and leave running for approx.
30 minutes before use.
Table 1:
230V / 50 Hz, 230V / 60 Hz, 220V / 60 Hz, 120V / 60 Hz, 100V / 50 Hz
100V / 60 Hz
35 VA MAX
133 x 152 x 275 mm
2.7 kg
Fire Retardant ABS Plastic
5A BS1362 (UK ONLY)
T1AL 250 V
Class II, Double Insulated with Functional Earth
Type BF
IPX0 - No protection
Continuous
60 seconds total
12 seconds inflation
48 seconds deflation
30 - 60 mmHg (±10 mmHg)
Temperature Range
+10 °C to +40 °C
(+50 °F to +104 °F)
+10 °C to +40 °C
(+50 °F to +104 °F)
-20 °C to +50 °C
(-4 °F to +122 °F)
14
Relative Humidity
30% to 75%
(non-condensing)
20% to 95%
(non-condensing)
20% to 95%
(non-condensing)
Atmospheric Pressure
700 hPa to 1060 hPa
700 hPa to 1060 hPa
500 hPa to 1060 hPa

Publicité

Table des Matières
loading

Table des Matières