ELECTROMAGNETIC COMPATIBILITY

Electromagnetic Interference - Although this equipment conforms with the intent of the directive
2014/30/EU in relation to Electromagnetic Compatibility (EMC), all electrical equipment may produce
interference. If interference is suspected, move equipment away from sensitive devices or contact the
manufacturer.
Portable and mobile RF communications equipment can effect medical electrical equipment.
Radios, cell phones and similar devices may affect this equipment and should be kept at least
2 m (6.5 ft) away from the equipment.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information in the following tables.
The following tables document compliance levels and guidance from the IEC 60601-1-2:2014 (4th edition)
Standard, for the electromagnetic environment in which the Skin IQ 1000 MCM should be used in a clinical
environment.
Guidance and manufacturer's declaration - electromagnetic emissions
The Skin IQ 1000 MCM is intended for use in the electromagnetic environment specified below. The customer or user of the
Skin IQ 1000 MCM should assure that it is used in such an environment.
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Compliance
The Skin IQ 1000 MCM uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to cause any
Class B
This equipment is suitable for use in all establishments, including domestic
Class B
establishments and those directly connected to the public low voltage
power supply network that supplies buildings used for domestic purposes.
yes
Electromagnetic Environment
interference in nearby electronic equipment.