Product description
Hinged cleaning aid for cleaning the aerosol head with a built-in funnel
opening and receptacle for filling 5 ml backwash liquid. The connection cord
and the control unit of the eBase Controller are also required for operation.
Information on connecting the controller to the power supply and on opera-
tion can be found in the Instructions for Use of the eFlow
2
Intended use
The easycare cleaning aid can help to shorten the nebulisa-
tion time if the performance of the aerosol head becomes
less than optimal.
With easycare cleaning, the pores in the membrane of the
aerosol head are mechanically rinsed free by backwashing in
the opposite direction.
The expected life cycle of the easycare cleaning aid is
3 years.
3
Important information for use
Cleaning with the easycare cleaning aid is recommended if the nebulisation
time for the same fill volume of a given substance increases significantly.
Additional information can be found in the current version of the eFlow
Instructions for Use in the Chapter "Troubleshooting".