Setting up the MiniLink
The serial number is on the flat side of the transmitter,
as shown on the right. See your monitor user guide for
steps on entering this number.
Connecting to the sensor
The MiniLink must be fully charged before connecting it
to the sensor.
1
After the sensor is inserted (follow the instructions provided with your
sensor), wait 5 minutes before connecting the transmitter. Check for
bleeding. Make sure that bleeding has stopped (see Warnings page for
details), then connect the MiniLink to the sensor.
2
Touch the inserted sensor at back of assembly to prevent movement.
3
Hold the MiniLink as shown to line up the
two notches on both sides with the flexible
side arms of the sensor. The MiniLink flat
side with the label should face the skin.
4
Slide the MiniLink onto the sensor and
push it in firmly until the flexible side arms
of the sensor "click" into the notches on
both sides of the MiniLink. In the next 20
seconds the MiniLink will flash green for
about 10 seconds with a proper connection.
5
If the MiniLink light does not flash, disconnect it from the sensor, wait for
several seconds and then reconnect. If the MiniLink light still does not
flash, charge the MiniLink.
6
After the MiniLink light flashes green, use your monitor to start
communicating with the sensor and to start the sensor initialization (see
your monitor user guide for instructions).
7
After the MiniLink successfully sends signals to the monitor, you may
choose to put an occlusive dressing over the MiniLink and the sensor.
Bathing and swimming
After the MiniLink and sensor are connected, they form a water-tight seal to a
depth of 8 feet (2.4 meters) for up to 30 minutes. You can shower and swim
- 6 -
SN GT2XXXXXXM
MEDTRONIC MINIMED
MiniLink Transmitter
REF MMT-7703XX
SN GT2XXXXXXM
FCC D: OH27703
IPX8
IC: 3458-7703
1234
sensor