Rotary Evaporators
RE300 RE301 RE302
RE300P RE301P RE302P
3
Vacuum and water connections
Twist grip height adjustment handle
Figure 1 Assembly
Assembly and operating instructions

Before use

If the equipment is not used in the manner described in this
manual the protection provided by the equipment may be
impaired.
Bibby Scientific rotary evaporators are designed to operate under the
following conditions:
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude up to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply voltage fluctuations not
greater than ±10% of nominal
❖ Over voltage category 2 IEC 60364-4-443
❖ Pollution degree 2 IEC664
2
9
1
7
5
On/Off switch
Rotating speed control
2
4
Location & Services
In order to get the optimum performance from your Bibby Scientific
rotary evaporator careful consideration should be given to the
intended location. Please study the following notes before
commencing assembly.
Choose a convenient location, which will allow easy access to both
rotating and receiving flask.
In order to allow the glassware to be raised and lowered, a
minimum headroom of 850mm is required.
The location should have access to the following services:
Cold water supply capable of delivering 60l/hr is required for RE300,
RE300P, RE301, RE301P.
Electricity supply
Drain
Vacuum line (optional)
Glass angle control
Additional height
adjustment knob