Turbo function will be stopped and rotation speed will return
to original setting. A symbol
Turbo function is stopped. Before this symbol disappears,
turbo function can not be activated again.
13
Error codes
Any malfunction during operation may be identified by an error
message on the display.
Proceed as follows in such cases:
Switch off device using the power switch
Carry out corrective measures
Restart the device.
Error
Cause
Effect
code
Error 3 Inside temperature is
motor off - Switch off the device and
too high
Error 4 Motor can't run normal
motor off - Switch off the device.
(can't start or not reach
the target speed)
Error 47
Overload
motor off - Switch off the device
Error 48
Error 50 Tube is protected in
motor off - Check the tube condition
order to avoid overload
If the actions described fails to resolve the fault or another error
code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
appears to indicate that the
The drive unit is maintenance-free but not immune from wear. The
carbon brushes of the motor wear down over the time.
Only use cleansing agents which have been recommended by IKA.
Dirt
Dyes
Construction materials
Cosmetics
Foodstuffs
Fuels
Solution
For materials which are not listed, please request information from
IKA application support.
let it cool down.
Wear protective gloves during cleaning the instruments.
- Restart the device.
Electrical instruments may not be placed in the cleansing agent for
- Check the structure for
the purpose of cleaning.
possible blockades
Do not allow moisture to get into the instrument when cleaning.
Before using another than the recommended method for cleaning
- Decrease the speed setting
or decontamination, the user must ascertain with IKA that this
or the load
method does not destroy the instrument.
- Decrease the speed setting
Spare parts order:
or the load
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare part, see www.ika.com,
14
Maintenance
For cleaning disconnect the main plug.
Cleaning agent
isopropyl alcohol
water containing tenside/isopropyl alcohol
water containing tenside/isopropyl alcohol
water containing tenside
water containing tenside
spare parts diagram and spare parts list.