Set Working Parameters; Target Speed (Target Rpm); Timer Value (Timer Sec); Timer Function - IKA ULTRA -TURRAX Tube Drive P control Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
12
Setting working parameters
12.1

Target speed (target rpm)

Set "target rpm" within "rpm max" range in the working screen
by rotating the "Rotating/Pressing" knob (B). Adjustment magni-
tude is 10 rpm. In remote control mode via computer, adjusting
magnitude is 1 rpm.
• Setting "target rpm" in working screen
 Turn the "Rotating/Pressing" knob (B) to change the value.
 Press "Rotating/Pressing" knob (B) to start working.
 Rotation symbol ( ,,) indicates running.
12.2

Timer function

Set "Timer" within "timer max" range in the working screen by ro-
tating the "Rotating/Pressing" knob (B). Adjustment magnitude is 10
seconds. In remote control mode via computer, adjusting magnitude
is 1 second.
If the "Timer" is deactivated when the device is started, the device
will automatically count from 0 to 30 minutes and then the unit will
stop running.
• Setting "Timer" value (timer sec) in working screen
 Press "Timer" button (D) to activate the timer setting in the
working screen.
 Turn the "Rotating/Pressing" knob (B) to change the value.
 Press "Timer" button (D) again to complete the setting.
 Current values are stored.
12.3
Reverse (Rev) function
When setting "Rev" value with the "Rotating/Pressing" knob (B), the
value can be set from 10 to 60 seconds where adjustment magni-
tude is 1 second.
Note: If both "Rev" function and "Timer" function are activated,
the "Rev" value should be less than "Timer" value. Otherwise, the
reverse function will not work.
If the reverse function is activated, press "Rev" button (E) dur-
ing running, the reverse function will be deactivated. Press "Rev"
button (E) again, the reverse function will be reactivated and the
screen will indicate "Rev" value setting.
• Setting "Rev" timer (rev sec) in working screen
 Press "Rev" button (E) to activate the reverse setting in the
working screen.
 Turn the "Rotating/Pressing" knob (B) to change the value.
 Press "Rev" button (E) again to complete the setting.
 Current values are stored.
12.4

Turbo Function

The rotation speed can increase to 8000 rpm in several seconds by
pressing down „Turbo" button (F) during operating. For tube protec-
tion, Turbo function can only last 60 seconds at most if you press
down „Turbo" button (F). Several seconds later, Turbo function can
be activated again after symbol has disappeared.
• Activate and hold "Turbo" function in working screen:
 Press down „Turbo" button (F) over 2 seconds when the de-
vice is working, to activate Turbo function.
 The Turbo mark
appears and changes alternately to in-
dicate that the Turbo function is activated and rotation speed
will increase to 8000 rpm in a short time.
 Release "Turbo" button (F).