17
Technical Information
Inlet screen installation
prevent the rotor blades from being damaged by debris. The inlet
screen, however, does reduce the pumping speed as per following
table:
The inlet screen is fitted in the upper part of the pump, as shown in
the figure.
306/336
Reduction in pumping speed (%)
Ar
X3502-
68000
20%
X3502-
68001
N
2
18%
TwisTorr 84 FS User Manual / 87-901-043-01(B)
He
H
2
7%
5%