Error Messages
Info text
Error
(is displayed in case of fault
when the Info key is pressed)
ERR 95
SYSTEM ERROR
ERR 96
WRONG MODULE
ERR 97
NO MODULE INSTALLED
ERR 98
CONFIGURATION ERROR
ERR 99
DEVICE FAILURE
ERR 100 INVALID SPAN OUT1
ERR 101 INVALID SPAN OUT2
ERR 105 INVALID SPAN I-INPUT
ERR 106 INVALID CHANNEL SELECTION
ERR 108 OUT1 INVALID CORNER X/Y
ERR 109 OUT2 INVALID CORNER X/Y
ERR 110
CATION EXCHANGER CAPACITY
16
Problem
Possible causes
System error
Restart required.
If error still persists, send in the
device for repair.
Wrong module
Please have the module replaced
at the factory.
Supply voltage too low
or no module installed
Error in confi guration or
calibration data
Memory error in device program
Configuration or calibration data
defective; completely reconfig-
ure and recalibrate the device.
Error in factory settings
EEPROM or RAM defective
This error message only occurs
in the case of a total defect. The
device must be repaired and
recalibrated at the factory.
Span Out1 configuration error
Span Out2 configuration error
I-Input configuration error
Configuration error
OUT1 bilinear,
wrong characteristic
OUT2 bilinear,
wrong characteristic
Capacity of ion exchanger used
up – replace