Info text
(is displayed in case of
Error
fault when the Info key is
pressed)
ERR 10
CONDUCTANCE TOO HIGH Conductance value
(channel A)
ERR 40
(channel B)
ERR 11
CONDUCTIVITY RANGE
(channel A)
ERR 41
(channel B)
ERR 13
TEMPERATURE RANGE
(channel A)
ERR 43
(channel B)
ERR 15
SENSOCHECK
(channel A)
ERR 45
(channel B)
ERR 59
INVALID CALCULATION
ERR 60
OUTPUT LOAD
ERR 61
OUTPUT 1 TOO LOW
ERR 62
OUTPUT 1 TOO HIGH
ERR 63
OUTPUT 2 TOO LOW
ERR 64
OUTPUT 2 TOO HIGH
ERR 71
I-INPUT
ERR 72
FLOW TOO LOW
ERR 73
FLOW TOO HIGH
Error Messages
Problem
Possible causes
out of range: > 250 mS
Display range violation
Cond > 9999 μS/cm
< 0.1 kohm cm
Temperature range
violation
Sensocheck
Check cable
pH value cannot be calcu-
lated. Diff erence between
the measured conductivity
values value is too large.
Load error
Output current 1 < 3.8 mA
Output current 1 > 20.5 mA
Output current 2 < 3.8 mA
Output current 2 > 20.5 mA
Current input limit value
Flow too low
Flow too high