12.9.
To define the upper and lower movement limits
12.9.1.
This function, MANUAL MODE, allows the maximum limits that a patient can tolerate to be defined and
recorded at the beginning of a session.
Set up the patient and proceed as below:
Beginning
Switch the unit ON
To select
MANUAL MODE
To set the pain level when reached,
immediately press
To set the lower limit
To set the pain level when reached,
immediately press
To start the session with the new movement
limits
12.9.2.
This function, BYPASS MODE, allows the maximum limits that a patient can tolerate to be defined and
recorded, which allows you to work on increasing amplitude.
Beginning
The unit is running
To select
BYPASS Mode
To set the new pain level when reached,
immediately press
To set the lower limit
To set the new pain level when reached,
immediately press
Continue the session with the NEW movement
limits.
Comments:
• The values shown in the 'Display' column are examples.
Kinetec Maestra™
At the start of a session
Keys to
press
continuously press
continuously press
During the session
Keys to
press
continuously press
continuously press
Display
KINETEC
MAESTRA V01.0
KINETEC
PLEASE WAIT
PRONO/SUPINATION
-10 STOP
5 10
PRONO/SUPINATION
-10 MANUAL
15
PRONO/SUPINATION
-10 MANUAL
15
PRONO/SUPINATION
-10 MANUAL
5
PRONO/SUPINATION
-5
MANUAL
5 15
PRONO/SUPINATION
-5 RUN
8
15
Display
PRONO/SUPINATION
-5 RUN
8
15
PRONO/SUPINATION
-5 BYPASS
40
PRONO/SUPINATION
-5 BYPASS
40
PRONO/SUPINATION
-10 BYPASS
30
PRONO/SUPINATION
-30
BYPASS
30 40
PRONO/SUPINATION
-30 RUN
20
EN
Remarks
Check that the hand control is not
locked (See page 6).
The unit starts moving to the upper
limit of the movement.
10
The settings are controlled by you.
The new upper value limit of the
15
movement is recorded.
The unit starts moving to the lower
15
limit of the movement.
The new lower value limit of the
movement is recorded.
The angle display changes with the
current movement.
Remarks
Check that the hand control is not
locked (See page 6).
The unit exceeds the recorded upper
15
limit.
The new upper value limit of the
40
movement is recorded.
The unit starts moving to the lower
40
limit of the movement.
The new lower value limit of the
movement is recorded.
The angle display changes with the
40
current movement.
12 / 23