Guide and declaration of the manufacturer - electromagnetic immunity
SOPRO device is intended to be used in the electromagnetic environment specified below.
Immunity trial
Conducted RF
3 Vrms 150 kHz to 80 MHz
EN 61000-4-6
3 V/m
Radiated RF
80 MHz to 2.5 GHz
EN 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency band applies.
Note 2: These recommendations may not apply in every situation. Electromagnetic wave propagation is
modified by the absorption and reflection due to the structures, objects and persons.
8
S o p r o 7 1 7 F i r s t • U s e r ' s m a n u a l • 2 0 1 0 - 0 3 - 2 6
E N G L I S H
The user should make sure it is used in this environment.
CEI 60601
Compliance
Severity level
3V
3V/m
Electromagnetic environment
level
Portable and mobile RF communication
devices should not be used at a distance from
SOPRO device including the cables, lower than
the recommended separation distance,
calculated with the applicable formulas
depending on the emitter frequency.
Recommended separation distance
d = 1,16 √P
d = 1,16 √P 80 MHz to 800 MHz
d = 2,33 √P 800 MHz to 2.5 GHz
where
is the maximum rated output of the
transmitter in watts (W) by the transmitter
manufacturer and d the recommended
separation distance in metres (m).
The field levels emitted by the fixed RF
transmitters,
electromagnetic measurement of the site
should be lower than the compliance level in
each frequency band.
Interference may occur in the vicinity of the
devices bearing the following symbol:
Guide
determined
by
an
a
,