Cycle

Use

Users must have familiarized themselves with the previous paragraphs, relating to the environment and safety,
before proceeding.
Follow the security procedures :
-
Never attempt to force the lid opening.
-
Always ensure that the pressure shown by the vessel pressure gauge is 0 before attempting to
open the lid.
-
Always manually operate the vessel's safety valve before attempting to open the lid, or the
adding or decanting port.
-
Should you need to move the device between 2 workstations, it is important to move it slowly
and to always hold it in both hands during the move.
-
Always put on protective gloves designed for high temperatures before accessing the vessel
→ certain surfaces can actually still be very hot even though the temperature of the medium
is below 80°C.
-
Never attempt to force the drain access hatch open.

Cycle screen

STAND-BY cycle
This screen appears when the device is started in stand-by mode. The last program used is suggested when
the device is switched on. Its no., name and a summary of the program's parameters are displayed.
The temperature of the vessel probe is displayed in real time in the left-hand column.