Step Start Alarm
Alarm 'Close Vessel Lid'
Alarm 'Close Safety Covers'
Heating error 'Abnormal
temperature variation'
Cooling error 'Abnormal
temperature variation'
Temperature warning '
temperature is out of tolerance'
Restart cycle after power on
'Media' temperature probe error
'Abnormal temperature
measurement'
Technical Error 'Stirrer motor'
Technical Error 'Pressure release
solenoid'
Technical Error 'Cooling'
Technical Error 'Locker magnet'
Communication Initialisation
failed
Communication failed. Please
switch off and contact the after-
sales service
Icônes
-
Follow the instruction of the message that have been created for this
message. Example: "Add blood"
The lid is not properly closed. Release the pressure from the vessel by
using the safety while the temperature is below 80°C. Then open the lid
and close it again to get blue light coming one. If not →Call service.
One of the safety lids (door trap or lid to access to the dispensing port)
is not closed or not properly closed--> Close the lids
Time to reach the sterilization set point is too long or the temperature is
not stable --> Make sure that the stirrer is running properly. If not, check
the bearing state.
Time to reach the dispensing set point is too long or the temperature is
not stable --> Make sure that the water tap is open.
The temperature was not in +/-1°C range during the sterilization plateau-
Make sure that temperature is within the tolerance of your application
before validating the cycle.
Just for information - Check the data records at the end of the cycle
before validating the cycle.
The vessel temperature probe is defective --> Call service
Defective stirrer --> Call service
Defective pressure release valve --> Call service
Defective cooling valve --> Call service
Defective lid locking solenoid --> Call service
Switch off and on again the machine - Call service if it does not solve the
problem
Switch off and on again the machine - Call service if it does not solve the
problem
Communication between HMI and automaton cards normal
Failure of communication between HMI and automaton cards normal
SD card connected and compliant
SD card failure
Connected to an Ethernet network
USB key connected