Easy Calibration

The calibration mode 'CAL'
Adjustment
The instrument should be set to either the nominal volume (for example
100 µl for a 100 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 40 for procedures.
What to do
1.
Bring up the CAL
mode
2.
Activate the CAL
mode
3.
Confirm CAL mode
4.
Set the volume
5.
Confirm volume
5 6
Adjustment of the Transferpette
TM
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display
reads 'off'. 'CAL' blinks.
Press one of the arrow
buttons to activate the CAL
mode. The display changes
from 'off' to 'on'. 'CAL'
continues to blink.
Press the ENTER key.
The display now shows
the set pipetting volume.
'CAL' blinks.
Use the arrow keys (+/-)
to set the volume, which
was previously determined
and tested. 'CAL' blinks.
Press the ENTER key.
The display shows the
tested and corrected
volume. The CAL symbol
is continously displayed to
confirm that an adjustment
has been made.
Keys to press
1x
>3s
1x
1x
®
-8/-12 electronic