Property
M12 connector
Sanitary clamp 2.5" / 4"
Cable
Weight
1)
Material certificate included.
2)
Manufacturer's declaration included.
3)
ADI free, FDA 21 C.F.R 177.1550, 3A Sanitary Standard, USP Class VI <88>, 70 °C.
4)
3-A certificate, EHEDG certificate.
Electrical specifications
Table 5 Inputs and outputs
Property
Supply
Operating voltage
Power consumption
Protection class
Outputs
Output parameters
Diagnostics
Analog outputs
mA
mA range
Specification
Gland, EN 1.4305 (AISI 303)
Contacts, CuZn with Ni/Au plating
Phoenix Contact, 1405233, M12/4(M), A,
2
4×0.34 mm
, TPE, 0.5 m
Carrier, PA 6.6
2)
EN 1.4301 (AISI 304)
4×22 AWG PUR, black 10 m multistrand, with
ferrules
Flame-retardant acc. to IEC 60332-1-2, FT1, VW1
3.6–4.2 kg (7.94–9.26 lb)
Specification
24 V DC nominal (9–30 V DC)
Less than 1 W
3, PELV
RI, temperature, concentration, quality factor
Image detection: outside light error, no optical
image, no sample, prism coated, prism coated
persistently, outside light to prism, low image
quality
Process temperature measurement: temperature
measurement fault
Concentration calculation: concentration out of
range, temperature out of range
Internal environment: high internal temperature,
high internal humidity
System: blank image corrupted
Sourcing, isolated, NAMUR NE 43, configurable
3.8–20.5 mA
27