• Install the refractometer indoors or outdoors. In outdoor installations, protect the device
from exposure to sunlight and rain.
• If the pipe is translucent, protect the prism from light. Outside light may disturb the
measurements.
• Use additional cooling according to the figure. Notice that the values in the figure are
indicative. Required cooling depends on the installation location and environment.
- 40
- 20
0
- 20
0
20
40
60
Installation dimensions for hygienic processes
When installing the refractometer in a hygienic, EHEDG compliant process, make sure the
length of the dead leg (L) is smaller than its inner diameter (D). With probe model
refractometers, follow the equation L≤(D-d).
1
Sensor
2
Dead space
3
Process area
Package recycling
Save the original packaging for safely transporting the device to Vaisala for servicing.