SYSTEM
GALVANIC:
13
Continuous
GALVANIC:
14
Interrupted
NO.
Waveform
TENS-
1
Asymmetrical
Biphasic
TENS- Symmetrical
2
Biphasic
3
VMS™
IFC (Interferential)
4
Traditional (4 Pole)
IFC Premodulated
5
(Traditional 2 Pole)
6
Russian
High Voltage Pulsed
7
Current (HVPC)
MONOPHASIC:
8
Monophasic
Rectangular pulsed
MONOPHASIC:
9
Monophasic
Rectangular pulsed
Diadynamic
10
Waveforms
11
Träbert (Ultrareiz)
12
Microcurrent
GALVANIC:
13
Continuous
GALVANIC:
14
Interrupted
NO.
Waveform
47979 INTELECT TRANSPORT 2 COMBO (4778) AND ULTRASOUND (4776) USER MANUAL
Use default settings, with polarity reversal and increase intensity to mild sensory
Pos; cont; 10 min
level/no pain
Pos; cont; 10 min
Use default settings and increase intensity to mild sensory level/no pain
Default settings
Note: Default settings
are the same for any
conditions. Please
adjust according to
this table
80Hz; cont; 20 min
80Hz; cont; 20 min
50Hz;2s ramp; cont;
20 min
4000Hz; 80/150Hz;
Use default settings and increase intensity to strong sensory
manual; cont; 20 min
level/no pain
2500Hz; 80/150Hz;
Use default settings and increase intensity to strong sensory
cont; 20 min
level/no pain
2500Hz; 50Hz; DC
50%; cont; 20 min
100Hz; cont; 20 min
Use default settings and increase intensity to strong sensory level/no pain
1.0ms/500ms; 5 min
1.0ms/500ms; 5 min
Diadyn settings (fixed),
Use default Diadynamic settings and increase intensity to strong sensory
10 min
level/no pain
Träbert settings (fixed),
Use default Träbert settings and increase intensity strong to sensory level/no pain
10 min
Neg; 1Hz; 20 min
Use default settings and increase intensity to 40-100µA
Pos; cont; 10 min
Pos; cont; 10 min
Default settings
Note: Default settings
are the same for any
conditions. Please
adjust according to
this table
Management and alleviation of post operative pain
Freq
Intensity
strong sensation/
80-150Hz
no pain
strong sensation/
80-150Hz
no pain
strong sensation/
80-150Hz
no pain
Muscle re-education
Freq
Intensity
EN
Duty cycle
Treatment Time
continuous
when in pain
continuous
when in pain
continuous
when in pain
when in pain
when in pain
when in pain
when in pain/
multiple hours
Duty cycle
Treatment Time

Publicité