Weighing Scale Settings Operating Mode Weighing; Filter - Adaptation To Ambient Conditions And Weighing Options - KERN and SOHN AEJ-C Série Mode D'emploi

Masquer les pouces Voir aussi pour AEJ-C Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12 Weighing scale settings operating mode weighing

This menu block can be used to adapt the behaviour of the weighing scale to your
requirements (such as ambient conditions, special weighing procedures) for any
operating mode.
Available settings for instance under menu item P2.2 WEIGHING

12.1 Filter - adaptation to ambient conditions and weighing options

Set the filter according to operating conditions for the weighing scale.
Filter setting
MEDIUM
SLOW – VERY SLOW
FAST – VERY FAST
Example: Menu item P2.2.1 invoke READING, press
(example)
AEJ-C/AES-C/PLJ-C-BA-e-1513
Balance
Normal ambient conditions
The balance reacts slowly and in a robust
manner - busy set-up location
Weighing scale reacts sensitively and fast
- Quiet place of installation
NORMAL
will
be
between settings, using the arrow keys
Save your selected setting by pressing
PRINT
You have successfully saved your filter
setting and the weighing scale will return
to menu.
To return to weighing mode, press UNIT
repeatedly.
flashing;
change
49

Publicité

Table des Matières
loading

Table des Matières